Presence and reliable amount are different claims
Detection vs quantitation limits distinguishes recognizing an analytical signal from measuring an amount with suitable performance. ICH Q2(R2) discusses lower-range limits as part of analytical validation. A detection threshold addresses whether a signal can be distinguished; a quantitation threshold addresses measurement at a lower level. Read the laboratory definition and reporting convention rather than assuming every number near the bottom of a report has the same evidential strength. [1]
Not detected does not mean absent
A not-detected statement is conditional on the method, sample preparation, and stated reporting threshold. It should not be rewritten as proof that the substance is entirely absent. An informative report supplies the relevant limit and units. If those are missing, the reader cannot tell how sensitive the investigation was. This distinction is especially useful when comparing certificates: two laboratories can use the same phrase while operating at different thresholds or examining different portions of a sample.
A detected signal may remain unquantified
A laboratory may recognize evidence of an analyte without reporting a reliable numerical amount. That is not automatically contradictory. The appropriate summary preserves the distinction between detection and quantification. Avoid manufacturing a precise value from an instrument screenshot or replacing a below-limit result with zero in a comparison table. If a data analysis requires a numerical convention for such results, that convention should be declared separately from the laboratory observation it represents.
Interpret the limit in context
Place the analyte, sample matrix, units, method, detection limit, quantitation limit, and reporting rule beside each other. Then ask whether the analytical sensitivity addresses the actual question posed by the study. A very small threshold is not sufficient if the method measures the wrong target or lacks suitable selectivity. Conversely, an appropriately validated method can answer a defined question without claiming unlimited sensitivity. These limits describe analytical capability; they are not stand-alone judgments about clinical safety or product acceptability.
Sources & reading notes
Source documentation checked 2026-11-25. A source link supports the nearby statement; it is not an endorsement of other claims on the source website.
- FDA ICH Q2(R2): validation of analytical proceduresAnalytical guidance